Highlights & Features
– 2"x3"x1.1" Package with base
– Up to 18.18W/inch3 power density
– Up to 120W output with natural
convection cooling
– Up to 150W peak power for 10 sec
– Up to 70°C Ambient Operation
– Up to 500KHrs MTBF
– 2MOPP Isolation
– Suited for Type BF Medical Products.
– Conformal Coating Option
– Class B Conducted and Radiated EMI
– IEC60601-1-2 4th Edition Immunity
Compliance
– For Class I/II system
Input |
|
|
|
|
Input Voltage | 90 Vac ~ 264 Vac |
|
|
|
Input Frequency | 47Hz ~ 63Hz |
|
|
|
Input Current | <1.3A |
|
|
|
Inrush Current | <60A cold start |
|
|
|
Power Factor | 0.98 typ. @ full load, 115Vac |
|
|
|
Efficiency | Up to 94% 1) |
|
|
|
Patient Leakage Current | <100uA normal, <500uA SFC |
|
|
|
Earth Leakage Current | <100uA normal, <300uA SFC |
|
|
|
Output |
|
|
|
|
Output Voltage | 12V | 15V | 18V | 24V |
Output Current (convection) | 0-9.5A | 0-8A | 0-6.67A | 0-5A |
No load Consumption | 0.21W |
|
|
|
Ripple & Noise | <1% Vrated pk-pk |
|
|
|
Environmental |
|
|
|
|
MTBF | 700KHrs |
|
|
|
Operation Temperature | -30°C ~70°C 1) |
|
|
|
Operation Altitude | 5000m or 54kPa |
|
|
|
1) Power de-rating with temperature above 50°C, refer to power de-rating curve for detail
Safety Certifications
- IEC60601-1 2nd edition
- IEC60601-1 3rd edition + A1 CB report
- TUV EN60601-1:2006/A11/A12
- UL60601-1+CAN/CSA 60601-1:
(Ed.3.2005)
- IEC60950-1 CB report
- IEC62368-1 CB report
- IEC60335-1 CB report (12V & 24V)
- IEC61558-1/-2-17 CB report (12V & 24V)
- GB4943.1-2011, GB9254-2008,
- GB17625.1-2012 (TBD)
详细规格请联系敝司销售人员